Volumetric infusion pump AP31
Simple, safe & reliable
Designed to meet the needs of modern I.V. therapy in
neonatal/pediatric/adult ICU, CCU, anesthesia, cardiology and oncology.
Key features:
user friendly software simplifies set-up
and operation
large, easily readable LCD display
wide range of infusion rate
drug dose calculation
automatic & manual bolus capability
easy modification of infusion parameters in
running course
comprehensive system of audible and visual
alarms & warnings
drug library with user selectable drug
names
RS232 interface
long battery life
data lock with a software-controlled
passkey system
event log (up to 2000 events recorded in
real time): on-screen display or as XML file
4 modes of operation: flow rate, flow rate
and volume, flow rate and time, volume over time
|
Performances |
|
|
Pump type: |
volumetric infusion pump |
|
Compatible I.V. set: |
DEHP/PVC-free UNI-ASCOSET infusion set |
|
Flow rate range: |
1 - 1000 ml/h
(programmable in 0.1 ml/h steps) |
|
Volume to be infused:
|
9 999,9 ml (programmable
in 0.1 ml steps) |
|
Maximum bolus rate:
|
up to 1500 ml/h
(programmable in 1 ml/h steps) |
|
Mass units
programming: |
units available:
ml/h, µg/h, mg/h, µg/kg/h, mg/kg/h, µg/kg/min, mg/kg/min |
|
Infusion time
selection: |
up to 9h 59m 59s |
|
KVO rate: |
0.1 - 5 ml/h
(programmable in 0.1 ml/h steps) |
|
Flow rate accuracy: |
± 5% |
|
Occlusion pressure
limit: |
40 - 80 kPa (300 - 600 mmHg)
- programmable in 1 kPa (75 mmHg) steps |
|
Air bubble detector: |
ultrasonic, alarm with
air bubbles of 0.01 - 0.05 ml in size or total of 1 ml accumulated over
the last 15 minutes of infusion |
|
Drop detector: |
optical, with digital
disturbance filtration |
|
Extended system of
alarms and warnings (both audible and visual): |
Alarms: occlusion,
end of infusion, KVO infusion, air in line, door open, no mains,
discharged battery, internal malfunction
Prealarms: 5 minutes left to the end of infusion, low battery |
|
|
>>back
|
|
Technical
specifications: |
|
|
Pumping system: |
peristaltic finger |
|
Power supply:
|
210 - 240 V AC, 50/60 Hz or 10 - 15 V DC |
|
Power consumption: |
max. 10 VA |
|
Build-in battery: |
rechargeable, NiMh 1300 mAh |
|
Battery life: |
min. 12 h at a flow rate of 25 ml/h
min. 3 h at a flow rate of 100 ml/h |
|
Battery charging
time: |
24 h |
|
Electrical safety: |
Class I, type BF,
IP31 |
|
Conformity standards:
|
EN 60601-1
EN 60601-1-2 (EMC)
EN 60601-2-24
MDD 93/42/EEC: IIb |
|
Weight: |
<3,4 kg |
|
Dimensions (WxDxH): |
140 x 230 x 230 mm |
|
Operating
temperature: |
5 - 40° C
relative humidity: 20 - 90% |
|
|
>>back
|
|
Other
features: |
|
|
Computer Interface: |
RS 232 C |
|
External alarm socket:
|
24 V, 1 A |
|
Other functions: |
battery life indicator
user selectable drug names infusion pressure
monitoring programmable name of ward, infusion parameters protected
by a password priming STAND-BY extended menu of user tests
|
|
Event log:
|
2000 events |
|
|
>>back
|
|
|
|