Single-syringe pump AP14
Simple, safe & reliable
Designed to meet the needs of modern I.V. therapy in neonatal/pediatric/adult ICU, CCU, anesthesia,
cardiology and oncology.
Key features:
user user friendly software simplifies set-up and operation
large & easily readable LCD display
wide range of infusion rate
over 55 types of compatiable syringes (5-60
ml)
drug dose calculation
automatic & manual bolus capability
automatic reduction of residual bolus after
occlusion release
easy modification of infusion parameters in
running course
comprehensive system of audible and visual
alarms & warnings
drug library with user selectable drug
names
RS 232C interface
long battery life
data lock with a software-controlled
passkey system
event log (up to 2000 events recorded in
real time): on-screen display or as XML file
4 modes of operation: flow rate, flow rate
and volume, flow rate and time, volume over time
|
Performances |
|
|
Flow rate range: |
0.1 - 2000 ml/h for
50 ml syringe
0.1 - 1200 ml/h for 30 ml syringe
0.1 - 1000 ml/h for 20 ml syringe
0.1 - 600 ml/h for 10 ml syringe
0.1 - 400 ml/h for 5 ml syringe
(programmable in 0.1 ml/h steps) |
|
Volume to be infused:
|
0.1 - 999 ml,
programable in 0.1 ml steps
|
|
Maximum bolus rate:
|
2000 ml/h for 50 ml
syringe
1200 ml/h for 30 ml syringe
1000 ml/h for 20 ml syringe
600 ml/h for 10 ml syringe
400 ml/h for 5 ml syringe |
|
Bolus dose /
Induction dose:
|
limited to syringe
size |
|
Mass units
programming: |
units available:
ml/h, µg/h, mg/h, µg/kg/h, mg/kg/h, µg/kg/min, mg/kg/min |
|
Drug concentration: |
0.01 µg/ml ÷ 9999
mg/ml |
|
Patient mass: |
up to 300 kg,
programable in 0.1 ml/h steps
|
|
Infusion time: |
up to 99 h 59 min 59
sec
|
|
KVO rate: |
0.1 - 5 ml/h,
programmable in 0.1 ml/h steps |
|
Flow rate accuracy: |
± 2% (complies with
EN 60601-2-24) |
|
Occlusion pressure
limit: |
9 - levels user selectable: 40 ÷ 120 kPa (300 ÷ 900 mmHg),
programmable in 10 kPa (75 mmHg) steps |
|
Compatible syringes: |
5/6, 10, 20, 30, 50/60 ml |
|
Syringe brands: |
B.BRAUN, BD, MONOJECT, NIPRO, CODAN, TERUMO, IVAC, INJECTOMAT,
PENTAFERTE and other (see operating manual) |
|
Extended system of
alarms and warnings
(both audible and visual): |
Alarms: no syringe, wrong syringe, empty syringe, end of infusion,
occlusion, no mains, discharged battery, internal malfunction.
Prealarms: 5 minutes left to the end of infusion or till the syringe gets empty,
low battery |
|
|
>>back
|
|
Technical
specifications: |
|
|
Power supply:
|
100 - 240 V AC, 50/60 Hz or from a built in rechageable battery,
optional 12 - 15 V DC for transport applications |
|
Power consumption:
|
max. 10 VA |
|
Built-in battery: |
rechargeable, NiMH 1300 mAh, automatically charged when the pump is
connected to AC power supply
|
|
Battery life: |
24 h at flow rate of 5 ml/h
4 h at flow rate of 100 ml/h |
|
Battery charging
time: |
24 h |
|
Classification: |
Class I, type CF, defibrillation proof, IP42 |
|
Conformity standards:
|
EN 60601-1
EN 60601-1-2 (EMC)
EN 60601-2-24
MDD 93/42/EEC:IIb |
|
Pump mass: |
2,5 kg |
|
Dimensions (W x H x
D): |
320 x 150 x 180 mm |
|
Operating conditions:
|
5° - 40° C
relative humidity: 20 - 90% |
|
|
>>back
|
|
Other
features: |
|
|
Computer Interface: |
RS 232C |
|
External alarm
socket: |
24 V, 1 A |
|
Other functions: |
anti-bolus system (ABS)
automatic syringe size recognition
battery life indicator
multiphase infusion (PROFILE mode)
user adjustable drug library
programmable name of ward
infusion pressure monitoring
Infusion parameters protected by a
password
priming
STAND-BY (24h)
extended menu of user tests and
service tests
|
|
Event log: |
2000 events |
|
|
>>back
|
|
|
|